Objective
Atrial fibrillation is the most common arrhythmia of clinical significance and hypertension is one of its major risk factors. This study aimed to evaluate the diagnostic accuracy of an automated oscillometric blood pressure (BP) monitor with the function of atrial fibrillation detection for atrial fibrillation screening.
Materials and methods
Patients attending outpatient cardiology clinics were recruited for atrial fibrillation screening by the BP monitor with triplicate BP measurements for atrial fibrillation detection. Furthermore, a single-lead ECG was recorded simultaneously for comparison as the reference standard. The diagnostic test’s evaluation index were analyzed, including sensitivity, specificity, and receiver operator characteristic (ROC) analysis.
Results
A total of 295 participants were analyzed including 166 males and 129 females, with an average age of 72.5 ± 5.9 years. The sensitivity and specificity for atrial fibrillation detection by the device were 1.000 and 0.904, respectively, with the area under the ROC curve of 0.952 (95% confidence interval: 0.929–0.975, P < 0.001). Furthermore, the device had a Kappa-value of 0.781 (P < 0.001) with the single-lead ECG in detecting atrial fibrillation.
Conclusion
The automated oscillometric BP monitor (G.LAB MD41A0) with atrial fibrillation detection function has high sensitivity and specificity with good accuracy for atrial fibrillation screening, which could be used as a reliable screening tool for the early detection of atrial fibrillation with potential benefits.
The Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors fulfilled the European Society of Hypertension revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols, and can be recommended for self-measurement in the general population.
The Grandway MD2301 digital automatic blood pressure monitor meets the standards of the ESH-IP revision 2010 and can be recommended for self/home measurement in the general population.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.