Background: Recombinant adeno-associated virus vectors based on serotype 2 (AAV-2) have become leading vehicles for gene therapy. Most humans in the general population have anti-AAV-2 antibodies as a result of naturally acquired infections. Pre-existing immunity to AAV-2 might affect the functional and safety consequences of AAV-2 vector-mediated gene transfer in clinical applications. Methods: An enzyme-linked immunosorbent assay (ELISA) method was developed using microwell plates coated with intact particles of recombinant AAV-2 vectors, and horseradish peroxidase-conjugated anti-human immunoglobulin G (HRP-IgG). Neutralizing antibody titres were analysed by assessing the ability of serum antibody to inhibit transduction into HEK293 cells of AAV vectors that express b-galactosidase. Results: Anti-AAV-2 antibodies were detected by ELISA in two of 20 healthy subjects. The positivity criterion (optical density .0.5) in ELISA corresponded to the cut-off value (320-fold dilution of serum) in the AAV-2 neutralization assay. Influences of interfering substances were not observed. Conclusion: This ELISA method may be useful for rapid screening of anti-AAV-2 neutralizing antibodies in candidates for gene therapy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.