Eight different recently synthesized macrocyclic diamides were studied to characterize their abilities as strontium ion carriers in PVC membrane electrodes. The electrode based on 1,13-diaza-2,3;11,12-dibenzo-4,7,10-trioxacyclopentadecane-14,15-dione exhibits a Nernstian response for Sr(2+) ions over a wide concentration range (1.0 × 10(-)(1)-3.2 × 10(-)(5) M) with a limit of detection of 8.0 × 10(-)(6) M (0.7 ppm). The response time of the sensor is ∼10 s, and the membrane can be used for more than three months without observing any deviation. The electrode revealed comparatively good selectivities with respect to many alkali, alkaline earth, and transition metal ions. It was used as an indicator electrode in potentiometric titration of carbonate ions with a strontium ion solution.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.