Decreased stride length is a highly relevant characteristic of the gait in patients with Parkinson's disease (PD). In this paper, a novel wearable laser device for stride length regulation is presented. The device is mounted to one foot and can project a red laser light strip to the floor as a visual cue for the other foot. In the experiment twelve healthy volunteers walked a 20 m straight walkway wearing the system on both feet. As an objective result, the stride length regulation reached an accuracy of 96.1 ± 2.5 (94.0 ± 3.5) % for a pre-defined stride length λdef= 55 (65) cm. The subjective evaluation by the participants using a questionnaire revealed that the visual cue projected from the laser device was considered a stable signal that did not shake during walking. 6 of 12 participants felt that this device was not interfering with their gait, 3 of 12 judged this aspect neutrally, and 3 of 12 considered it somehow bothersome.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.