The overall uncertainty associated with the use of XR-RV3 films to determine skin dose in the interventional environment can realistically be estimated to be around 20% (k = 1). This uncertainty can be reduced to within 5% if carefully monitoring scanner, film, and fitting-related errors or it can easily increase to over 40% if minimal care is not taken. This work demonstrates the importance of appropriate calibration, reading, fitting, and other film-related and scan-related processes, which will help improve the accuracy of skin dose measurements in interventional procedures.
The use of plane parallel ionization chambers with electron beams with high dose per pulse entails dose uncertainties due to the overestimation of the ion recombination factor, k, up to 20% if conventional dosimetric protocols are used. In this work MD-55-2 radiochromic films have been used as reference dosimeters to obtain dose to water per pulse DGAF(w) values for three Novac7 (Hitesys) electron beams of E0 = 5.8 MeV. However, the beam calibration by MD-55-2 films is time consuming and the use of plane parallel chambers is fundamental for a periodic quality control procedure. Three plane parallel chambers have been used and the general formula for the k determination has been tested using the calibration doses, DGAF(w). In particular, consistent ion recombination factors ksat(V0) (with the ion chamber polarized at V0), that follow the Boag theory, have been estimated at different dose per pulse values for the three plane parallel ionization chambers. This means that at present any ion chamber needs a specific ksat (V0) determination by using a reference dosimeter for which the response is independent of the dose rate. An accurate determination of ksat(V0), using a reference quality beam, can be used to determine the dose to water per pulse for electron beams of different quality and geometrical configuration.
In this work some dosimetric characteristics of MD-55-2 GafChromic films were studied in a low energy proton beam (21.5 MeV) directly in a water phantom. The nonlinearity of the optical density was quantified by a factor P(lin). A correction factor P(en), that accounts for optical density dependence on the energy, was empirically determined. The effects of detector thickness in depth dose measurements and of the film orientation with respect to beam direction were investigated. The results show that the MD-55-2 films provide dose measurements with the films positioned perpendicularly to the proton beam. A dosimetric formalizm is proposed to determine the dose to water at depth d, with films oriented perpendicularly to the beam axis. This formalism uses a calibration factor of the radiochromic film determined directly on the proton beam at a reference depth in water, and the P(lin) factor, that takes into account the nonlinearity of the calibration curve and the P(en) factor that, in turn takes into account the change of proton beam energy in water. The MD-55-2 films with their high spatial resolution and the quasiwater equivalent material are attractive, positioned perpendicularly along the beam axis, for the absolute dose determination of very small beam sizes and modulated proton beams.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.