A current challenge for development of amperometric sensors for the greenhouse gas nitrous oxide (NO) is their sensitivity toward dioxygen and trace water. The need for aqueous dioxygen scavengers in front of the sensor implies a background signal from penetrating water vapor. In this paper, we introduce substituted phosphines as dioxygen scavengers and demonstrate the application in a dioxygen-insensitive NO sensor. Suitably substituted phosphines have been synthesized to achieve good solubility properties in the electrochemically inert solvent propylene carbonate. Several sensors with and without physical separation of the sensing and dioxygen scavenging compartments were made and compared to current commercial sensors. The use of phosphines soluble in organic solvents as dioxygen scavengers yielded a higher sensitivity, albeit with longer response time. Proof-of-concept NO sensors without the physically separated dioxygen scavenger chamber showed a greatly enhanced sensitivity with a comparable response time, thus demonstrating the possibility for greatly simplified sensor construction.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.