Subacute stent thrombosis, caused by undesired interactions between blood and the stent surface, is a major concern in the first few weeks following coronary artery stent implantation. The aim of this study was to establish a novel in vitro model for hemocompatibility testing of coronary artery stents according to ISO 10993-4. The model consists of a modified Chandler-Loop design with closed heparin-coated PVC Loops and a thermostated water bath. The tests were performed with anticoagulated human whole blood. After incubation in the loop, blood was analyzed for coagulation and inflammatory activation markers (TAT, β-TG, sP-selectin, SC5b-9 and PMN-elastase). Three different stent types with varying thrombogenicity were tested; statistically significant differences were found between the three stent types in measures of coagulation and platelet activation. The new Chandler-Loop model can be used as an alternative to animal and current in vitro models, especially for the determination of early events after stent implantation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.