Testosterone (T) and its metabolite dihydrotestosterone (DHT) are androgens with different biologic profiles. T and DHT measurements are required for assessment of patients with ambiguous genitalia, hirsutism, during 5 alpha reductase treatment of prostate disorders, and new androgen formulations. Our laboratory has developed and validated a method to simultaneously measure serum T and DHT with liquid chromatography tandem mass spectrometry (LC-MS/MS) for use in a clinical chemistry laboratory. Analysis of sera from blood collected in tubes containing clot activator gave results of T that were fourfold higher than blood collected in plain tubes. Changing the ion pair selected for monitoring eliminated this interference by clot activators. Blood collected in fluoride-coated tubes gave serum T and DHT levels that were 20 and 15% lower, respectively than levels measured in blood collected in plain tubes (no additives). Addition of T enanthate to blood collected in plain tubes caused a dose related increase serum T levels due to the action of non-specific esterases in the red cells. This esterase activity could be avoided by using fluoride tubes for blood collection. Serum DHT levels were consistently lower when measured by LC-MS/MS versus radioimmunoassay. The differences were concentration dependent and the variance for the difference was large when serum DHT concentration was low. Celite chromatograph prior to radioimmunoassay reduced the differences between the two methods, thus confirming that higher levels of DHT obtained by immunoassays were probably due to interfering substances which were partially removed by Celite chromatography.
Testosterone (T) and its metabolite dihydrotestosterone (DHT) are
androgens with different biologic profiles. T and DHT measurements are required
for assessment of patients with ambiguous genitalia, hirsutism, during 5 alpha
reductase treatment of prostate disorders, and new androgen formulations. Our
laboratory has developed and validated a method to simultaneously measure serum
T and DHT with liquid chromatography tandem mass spectrometry (LC-MS/MS) for use
in a clinical chemistry laboratory. Analysis of sera from blood collected in
tubes containing clot activator gave results of T that were four fold higher
than blood collected in plain tubes. Changing the ion pair selected for
monitoring eliminated this interference by clot activators. Blood collected in
fluoride coated tubes gave serum T and DHT levels that were 20 and 15 percent
lower respectively than levels measured in blood collected in plain tubes (no
additives). Addition of T enanthate to blood collected in plain tubes caused a
dose related increase serum T levels due to the action of non-specific esterases
in the red cells. This esterase activity could be avoided by using fluoride
tubes for blood collection. Serum DHT levels were consistently lower when
measured by LC-MS/MS versus radioimmunoassay. The differences were concentration
dependent and the variance for the difference was large when serum DHT
concentration was low. Celite chromatograph prior to radioimmunoassay reduced
the differences between the two methods, thus confirming that higher levels of
DHT obtained by immunoassays were probably due to interfering substances which
were partially removed by Celite chromatography.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.