Flow-injection (FI) incorporated with inductively coupled plasma mass spectrometry (ICP-MS) was used to determine lead in aqueous solutions and seawater. Lead was retained in the form of Pb-DDPA complex in a sorbent microcolumn packed with C18-bonded silica, and then carried to an ultrasonic nebulizer with methanol as eluent. The ultrasonic nebulizer removed about 66% of methanol solvent vapors from entering the detector. A multi-tube Nafion® membrane dryer, installed along the passage prior to the ICP torch, further removed 50% of methanol. Reducing the amount of methanol solvent significantly improved the signal-to-background ratio of the detector. Accordingly, when the sample was loaded for 15 and 60 s, respectively, Pb detection limits were determined to be 0.162 and 0.028 ng/L, one to two orders of magnitude lower than those reported. The enrichment factors were also measured to be 7 and 28, with the sampling frequencies of 31 and 21 h−1, respectively. The method was applied for the determination of Pb in a reference seawater. The obtained result was 0.013 ± 0.001 ng/mL, consistent with the certified value of 0.013 ± 0.005 ng/mL.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.