This report describes the details of a recently upgraded spectral irradiance measurement facility in optical radiation standards at National Physical Laboratory, India. This facility provides the calibration of spectral irradiance in the wavelength range 280 nm -2500 nm. PTB, Braunschweig, Germany calibrated five numbers of 1000 W quartz halogen lamps, which are used as reference standards for spectral irradiance scale. In addition to providing the details of instruments, the procedure of calibration and evaluation of uncertainties is also described. For checking the fidelity, repeatability and reproducibility of the upgraded system, calibration of one of the PTB calibrated lamps, was done against the other four PTB calibrated lamps. These measurements not only provide confidence on the upgraded system but also verify the retention of the PTB certificate values after a lapse of two years after their calibration.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.