In 2012, a plan to develop Stereotactic treatments using a Cyberknife was unveiled at the Hermitage Medical Clinic, Dublin. Due to planning restrictions the new facility had to be contained in the existing hospital's blue print with the only available location being an unused CT simulation room. The room design would be different from conventional radiotherapy bunkers due to the fact the Cyberknife can fire an unfiltered beam in any direction bar the roof (restriction of 22° above the horizontal). Therefore all walls must be primary barriers with the roof designed to protect against the large leakage radiation resulting from the high MU's used during the treatments. Space consideration indicated that concrete alone could not be used to restrict the radiation beam to acceptable limits. To this end a combination of steel, lead, normal and heavy concrete were used to meet the dose constraints established by the Irish licensing authorities.
(a) The add-on microMLC system in combination with our LINAC has been commissioned to be used for clinical purposes and showed good agreement with published results for different ,MLC types. (b) This work has lead to the recommendation that leaves should be recalibrated after ten static beams or after each dynamic arc.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.