High tissue contrast in magnetic resonance imaging (MRI) allows better radiotherapy planning. However, geometric distortion in MRI induces inaccuracies affecting such planning, making it necessary to evaluate the characteristics of such geometric distortion. Although many studies have considered geometric distortion, most of these involved measurements performed only a few times. In this study, we evaluated MRI device-specific geometric distortion over long term and measured its variation by using an automatic analysis tool. The result showed that geometric distortion increased with distance from the center along both lateral and longitudinal directions. Specifically, the average distortion rate and average diameter error over the full measurement period increased by up to 1.02% and 1.96 mm, respectively, when using T1 weighted Image (WI) 3D fast spoiled gradient echo (FSPGR) at R15. In the case of T2 WI 2D fast spin echo (FSE) at R15, the standard deviation of the distortion rate and diameter error increased up to 0.38%, 0.72 mm, respectively. We conclude that periodic quality assurance of geometric distortion should be performed in order to maintain geometric distortion within allowable values.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.