This paper describes the design of a new, versatile, and low‐cost on‐line LC‐GC interface that allows the fast and reliable introduction of large sample volumes onto a capillary GC column. The sample introduction procedure consists successively of: evaporation of the entire sample (LC fraction), selective removal of the solvent and simultaneously cold‐trapping of the solutes, splitless transfer of the solutes to the GC column, on‐column focusing, GC separation and detection. Quantitative and qualitative aspects of various experimental parameters are evaluated and optimum conditions are reported. The applicability of the method is demonstrated on a synthetic aqueous sample of chlorinated pesticides.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.