Quantitative analysis of red phosphorus in polypropylene was studied using a temperature programmable pyrolyzer in combination with a mass spectrometer. Evolved gas analysis (EGA) profiles are obtained by continuous measurements of evolved gases from a sample as heating the sample at a constant heating rate. During heating the sample, red phosphorus sublimates into P4 molecules which have characteristic ions (m/z 31, 62, 93 and 124). Red phosphorus in polypropylene was determined from the m/z 62 ion peak area of the EGA profile with good reproducibility. The determined value was close to the value of original formulation and to the one determined by pyrolysis-GC/MS.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.