Objective: Our purpose was to evaluate the results of Vibrant Soundbridge (VSB) in conductive or mixed hearing loss. Materials and Methods: Twenty-five adult patients (29 ears) with a mixed or conductive hearing loss and various etiologies were included in this retrospective study. The preoperative ipsilateral pure tone average was 71 ± 3.0 dB, and the average bone conduction threshold was 42 ± 2.8 dB (n = 29). The transducer was placed on the long apophysis of the incus (n = 16), in the round window (n = 10) or on the stapes (n = 3). Results: No complications were noted. The bone conduction threshold remained unchanged. VSB was activated in all cases. The postoperative pure tone average without VSB was 63 ± 3.9 dB (n = 24) and with VSB in free-field condition 24 ± 2.1 dB (n = 22). Conclusions: VSB is safe and efficacious for auditory rehabilitation in conductive and mixed hearing losses.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.