Background/Aims: Access blood flow (Qa) measurements are recommended by the current guidelines as one of the most important components in vascular access maintenance programs. This study evaluates the efficiency of Qa measurement with on-line conductivity (OLC-Qa) and blood temperature monitoring (BTM-Qa) in comparison with the gold standard saline dilution method (SDM-Qa). Subjects and Methods: 50 long-term hemodialysis patients (42 arteriovenous fistulas/8 arteriovenous grafts) were studied. Bland-Altman and Lin’s coefficient (ρc) were used to study accuracy and precision. Results: Mean values were 1,021.7 ± 502.4 ml/min SDM-Qa, 832.8 ± 574.3 ml/min OLC-Qa (p = 0.007) and 1,094.9 ± 491.9 ml/min with BTM-Qa (p = NS). Biases and ρc obtained were –188.8 ml/min (ρc 0.58) OLC-Qa and 73.2 ml/min (ρc 0.89) BTM-Qa. The limits of agreement (bias ± 1.96 SD) obtained were from –1,119 to 741.3 ml/min (OLC-Qa) and –350.6 to 497.2 ml/min (BTM-Qa). Conclusions: BTM-Qa and OLC-Qa are valid noninvasive and practical methods to estimate Qa, although BTM-Qa was more accurate and had better concordance than OLC-Qa compared with SDM-Qa.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.