This paper describes the manufacturing process of quartz-pellets and shows their potential use as thermoluminescence dosimeters (TLD) for those applications where low-levels of ionising radiation are present. Two batches of cold-pressed pellets were produced and their resistance were evaluated by vibration tests and weight-loss measurements. The batch manufactured with 75 × 150 µm particles showed enough resistance to be employed as TLD. The dosimetric properties of the 310 °C peak appearing in the glow curves of these pellets were characterized together with commercial TLD-100 units using γ-and X-ray beams with different energies. The uncertainties related to reproducibility and stability of the TL signal were better than 10%. The sensitivity and the linearity of the TL response of quartz-pellets were better than that measured for TLD-100 for doses ranging from 0.5 to 200 mGy. The energy dependence of the quartz-pellets was higher than that of TLD-100 but it cannot be considered a restriction to their use in clinical procedures and industrial applications.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.