The purpose of the present study was to evaluate the interface shear strength of unloaded titanium implants with a sandblasted and acid-etched (SLA) surface in the maxilla of miniature pigs. The two best documented surfaces in implant dentistry, the machined and the titanium plasma-sprayed (TPS) surfaces served as controls. After 4, 8, and 12 weeks of healing, removal torque testing was performed to evaluate the interface shear strength of each implant type. The results revealed statistically significant differences between the machined and the two rough titanium surfaces (p <.00001). The machined surface demonstrated mean removal torque values (RTV) between 0.13 and 0.26 Nm, whereas the RTV of the two rough surfaces ranged between 1.14 and 1.56 Nm. At 4 weeks of healing, the SLA implants yielded a higher mean RTV than the TPS implants (1.39 vs. 1. 14 Nm) without reaching statistical significance. At 8 and 12 weeks of healing, the two rough surfaces showed similar mean RTVs. The implant position also had a significant influence on removal torques for each implant type primarily owing to differences in density in the periimplant bone structure. It can be concluded that the interface shear strength of titanium implants is significantly influenced by their surface characteristics, since the machined titanium surface demonstrated significantly lower RTV in the maxilla of miniature pigs compared with the TPS and SLA surfaces.
In recent years, nonfusion stabilization of the lumbar spine has gained more and more popularity. These nonfusion systems intend to maintain or restore the intersegmental motions to magnitudes of the intact spine and have no negative effects on the segments adjacent to the stabilized one. This study investigated the DYNESYS, a dynamic nonfusion system, which is designed to stabilize the bridged segments while maintaining the disc and the facet joints. To determine the magnitude of stabilization and the effect of the stabilization on the adjacent segment, six lumbar cadaver spines were fixed in a spine tester and loaded with pure moments in the three main motion planes. For each spine, four different stages were tested: intact, defect of the middle segment, fixation with the DYNESYS, and fixation with the internal fixator. Intersegmental motions were measured at all levels. For the bridged segment, the DYNESYS stabilized the spine and was more flexible than the internal fixator. This difference between the internal fixator and the DYNESYS was most pronounced in extension (P < 0.05), with the DYNESYS restoring the motion back to the level of the intact spine. The motion in the adjacent segments was not influenced by either stabilization method. Our results suggest that the DYNESYS provides substantial stability in case of degenerative spinal pathologies and can therefore be considered as an alternative method to fusion surgery in these indications while the motion segment is preserved.
Using nonmetallic CF/PEEK instead of standard titanium as pedicle screw material did not affect screw loosening in the chosen test setup, whereas cement augmentation enhanced screw anchorage of CF/PEEK screws. While comparable to titanium screws in terms of screw loosening, radiolucent CF/PEEK pedicle screws offer the significant advantage of not interfering with postoperative imaging and radiotherapy. These slides can be retrieved under Electronic Supplementary Material.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.