Objective: To evaluate the long-term safety and performance of four different vibroplasty couplers (round window, oval window, CliP and Bell coupler) in combination with an active middle ear implant. Methods: This was a multicentre, prospective, long-term study including 5 German hospitals. Thirty adult subjects suffering from conductive or mixed hearing loss were initially enrolled for the study, 24 of these were included in the final analysis with up to 36 months of postsurgical follow-up data. Bone conduction and air conduction were measured pre- and postoperatively to evalu ate safety. Postoperative aided sound field thresholds and Freiburger monosyllable word recognition scores were compared to unaided pre-implantation results to confirm performance. Additional speech tests compared postoperative unaided with aided results. To determine patient satisfaction, an established quality-of-life questionnaire developed for conventional hearing aid usage was administered to all subjects. Results: Mean postoperative bone conduction thresholds remained stable throughout the whole study period. Mean functional gain for all couplers investigated was 38.5 ± 11.4 dB HL (12 months) and 38.8 ± 12.5 dB HL (36 months). Mean word recognition scores at 65 dB SPL increased from 2.9% in the unaided by 64.2% to 67.1% in the aided situation. The mean postoperative speech reception in quiet (or 50% understanding of words in sentences) shows a speech intelligibility improvement at 36 months of 17.8 ± 12.4 dB SPL over the unaided condition. The signal-to-noise ratio (SNR) improved by 5.9 ± 7.2 dB SNR over the unaided condition. High subjective device satisfaction was reflected by the International Inventory for Hearing Aids scored very positively. Conclusion: A significant improvement was seen with all couplers, and audiological performance did not significantly differ between 12 and 36 months after surgery.
In patients with mixed hearing loss combined with a destroyed stapes suprastructure but a mobile stapes footplate, we describe the coupling of the floating mass transducer component of a Vibrant Soundbridge to the stapes footplate, as an alternative to coupling to the round window.
Objective: To evaluate the safety and effectiveness of round window (RW), oval window (OW), CliP and Bell couplers for use with an active middle ear implant. Methods: This is a multicenter, long-term, prospective trial with consecutive enrollment, involving 6 university hospitals in Germany. Bone conduction, air conduction, implant-aided warble-tone thresholds and Freiburger monosyllable word recognition scores were compared with unaided preimplantation results in 28 moderate-to-profound hearing-impaired patients after 12 months of follow-up. All patients had previously undergone failed reconstruction surgeries (up to 5 or more). In a subset of patients, additional speech tests at 12 months postoperatively were used to compare the aided with the unaided condition after implantation with the processor switched off. An established quality-of-life questionnaire for hearing aids was used to determine patient satisfaction. Results: Postoperative bone conduction remained stable. Mean functional gain for all couplers was 37 dB HL (RW = 42 dB, OW = 35 dB, Bell = 38 dB, CliP = 27 dB). The mean postoperative Freiburger monosyllable score was 71% at 65 dB SPL. The postimplantation mean SRT50 (speech reception in quiet for 50% understanding of words in sentences) improved on average by 23 dB over unaided testing and signal-to-noise ratios also improved in all patients. The International Outcome Inventory for Hearing Aids (IOI-HA)quality-of-life questionnaire was scored very positively by all patients. Conclusion: A significant improvement was seen with all couplers, and patients were satisfied with the device at 12 months postoperatively. These results demonstrate that an active implant is an advantage in achieving good hearing benefit in patients with prior failed reconstruction surgery.
During simulated tensor traction the aspect of the tympanic membrane changed slightly, i.e. there was some inward movement of the umbo. However, such effects were only identifiable during the pulling action or by directly comparing the "contracted" and "relaxed" states. Tympanometry revealed a decrease in the peak amplitudes and a shift in the middle ear resonance towards higher frequencies during contractions.
Intraoperative stapedius reflex recording by bipolar electromyography via a hook electrode is feasible. Further research will show whether these signals may be used for establishing a self-adjusting speech processor.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.