Standardization of rotational chair testing across laboratories has not been achieved because of differences in test protocol and analysis algorithms. The Interlaboratory Rotational Chair Study Group was formed to investigate these differences. Its first study demonstrated significant variability in calculated results using actual patient data files. No estimation of accuracy could be made, however, because the "true" values of response parameters were unknown. In this study we used simulated "patient" data files to further explore the differences among analysis algorithms. We found a high degree of agreement and accuracy across laboratories using automated analysis of high signal-to-noise/low-artifact data for gain, phase, and asymmetry. Variability increased significantly for the lower signal-to-noise ratio/higher artifact files. Operator intervention generally improved accuracy and decreased variability, but there were cases in which operator intervention reduced accuracy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.