Background: Diagnostic testing for chronic esophageal disorders rely on histopathology analysis of biopsies or uncomfortable transnasal catheters or wireless pH monitoring, which capture abnormal intraluminal refluxate. We therefore developed a balloon mucosal impedance (MI) catheter system that instantly detects changes in esophageal mucosal integrity during endoscopy over a long segment of the esophagus. We performed a prospective study to evaluate the ability of a balloon-incorporated MI catheter to detect and evaluate esophageal disorders, including gastroesophageal reflux disease (GERD) and eosinophilic esophagitis (EoE). Methods: We performed a prospective study of 69 patients undergoing esophagogastroduodenoscopy with or without wireless pH monitoring. Patients were classified as having GERD (erosive esophagitis or abnormal pH; n = 24), EoE (confirmed with pathology
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.