Arsenic, lead, and cadmium were determined in tap and bottled water samples consumed in the west part of Turkey at trace levels. Graphite furnace atomic absorption spectrometry (GFAAS) was used in all detections. All of the system parameters for each element were optimized to increase sensitivity. Pd-Mg mixture was selected as the best matrix modifier for As, while the highest signals were obtained for Pb and Cd in the case of Ni used as matrix modifier. Detection limits for As, Cd, and Pb were found to be 2.0, 0.036, and 0.25 ng/mL, respectively. 78 tap water and 17 different brands of bottled water samples were analyzed for their As, Cd, and Pb contents under the optimized conditions. In all water samples, concentration of cadmium was found to be lower than detection limits. Lead concentration in the samples analyzed varied between N.D. and 12.66 ± 0.68 ng/mL. The highest concentration of arsenic was determined as 11.54 ± 2.79 ng/mL. Accuracy of the methods was verified by using a certified reference material, namely, Trace Element in Water, 1643e. Results found for As, Cd, and Pb in reference materials were in satisfactory agreement with the certified values.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.