Background: Achieving a favorable pacing threshold with a Micra transcatheter pacing system (Micra-TPS) is needed to reduce battery depletion. In some cases, the threshold increases shortly after the device is implanted, and a higher pacing threshold may be required. This study aims to identify the causes and predictors of the increase in pacing threshold observed shortly after Micra-TPS implantation. Methods: The study included 64 consecutive patients who underwent Micra-TPS implantation between 2017 and 2020. The patients were divided into two groups depending on their pacing threshold: the increased pacing threshold (IPT) group (threshold increased by ≥0.5 V/0.24 ms within 1 month of implantation) and the stable pacing threshold (SPT) group. Results: Excluding four patients who could not be followed up, of the 60 remaining patients, nine (15%) were in the IPT group and 51 (85%) were in the SPT group. The IPT group had significantly lower implant impedance values and higher implant thresholds than the SPT group (582 ± 59 vs 755 ± 167 Ω [P < .001] and 1.29 ± 0.87 vs 0.71 ± 0.40 V/0.24 ms [P = .014]). Implant impedance and threshold may serve as predictors of a threshold increase after implantation (area under the curve: 0.737-0.943 and 0.586-0.926, respectively). Conclusions: An IPT was noted shortly after Micra-TPS implantation owing to microdislodgement because of insufficient anchoring of the device to the myocardium. Impedance >660 Ω and threshold <1.0 V/0.24 ms may predict an increase in pacing threshold.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.