Background
Automatic lancets have been reported to be superior to manual lancets in terms of pain and treatment time. However, no studies have yet been published comparing automatic lancet and needle puncture heel‐prick blood sampling. The objective of this study was to compare the pain response and efficiency between the automatic lancet and needle at the time of heel blood sampling. The design was a randomized controlled trial. The inclusion criteria for the participants were a birthweight of ≧1,500 g and a gestational age of ≧30 weeks.
Methods
The study examined a total of 105 neonates who were randomized into an automatic lancet group (n = 53) and a needle group (n = 52). The parameters measured included blood collection time, number of calf squeezes, duration of audible crying, and the Neonatal Infant Pain Scale (NIPS) score. The main outcome measure was audible crying duration.
Results
The duration of audible crying was significantly shorter in the automatic lancet group when compared to the needle group (median 3 s, interquartile range (IQR) 0–33 s vs median 39 s, IQR 5–91.5 s, P = 0.0023). The NIPS score at the time of puncture was significantly lower in the automatic lancet group than in the needle group (median 1, IQR 0–5 vs median 5, IQR 3–6, P = 0.0060). There was no significant difference in the blood collection time and the number of calf squeezes between the two groups. The automatic lancet was found to be less painful than the needle puncture in neonatal heel‐prick blood sampling with no significant difference in blood sampling time.
Conclusion
The automatic lancet was found to be less painful than the needle puncture in neonatal heel‐prick blood sampling with no significant difference in blood sampling time.
The purpose of this study was to determine the accuracy of mean blood flow velocity (mean V) in the internal carotid artery (ICA) for prediction of outcome in infants with hypoxic-ischemic encephalopathy (HIE) exposed to therapeutic hypothermia (TH). Five newborns with HIE who met the criteria for TH were enrolled. Ultrasonography of the right and left ICA was performed before, during, and after TH. Mean V of the sampling point in each ICA was measured. Mean V was suppressed during TH and increased after rewarming in four infants with normal neurological development. In one infant with neurological disability, however, mean V increased during TH and decreased after therapy. In conclusion, cervical ultrasonography for ICA in infants during TH may be useful for the prediction of neurodevelopmental outcome.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.