We compared images obtained using a three-dimensional iterative image reconstruction (3D-IIR) algorithm for C-arm-based interventional cone-beam computed tomography (CBCT) with that using the customary reconstruction technique to quantify the effect of reconstruction techniques on image quality.
We scanned 2 phantoms using an angiography unit with digital flat-panel system—an elliptical cylinder acrylic phantom to evaluate spatial resolution and a Catphan phantom to evaluate CT number linearity, image noise, and low-contrast resolution. Three-dimensional imaging was calculated using Feldkamp algorithms, and additional image sets were reconstructed using 3D-IIR at 5 settings (Sharp, Default, Soft+, Soft++, Soft+++). We evaluated quality of images obtained using the 6 reconstruction techniques and analyzed variance to test values of the 10% value of each MTF, mean CT number, and contrast-to-noise ratio (CNR), with
P
< .05 considered statistically significant.
Modulation transfer function curves and CT number linearity among images obtained using the customary technique and the 5 3D-IIR techniques showed excellent agreement. Noise power spectrum curves demonstrated uniform noise reduction across the spatial frequency in the iterative reconstruction, and CNR obtained using all but the Sharp 3D-IIR technique was significantly better than that using the customary reconstruction technique (Sharp,
P
= .1957; Default,
P
= .0042; others,
P
< .0001). Use of 3D-IIR, especially the Soft++ and Soft+++ settings, improved visualization of low-contrast targets.
Use of a 3D-IIR can significantly improve image noise and low-contrast resolution while maintaining spatial resolution in C-arm-based interventional CBCT, yielding higher quality images that may increase safety and efficacy in interventional radiology.
Objective: The purpose of this study was to evaluate the image quality in virtual monochromatic imaging (VMI) at 40 kilo-electron volts (keV) with three-dimensional iterative image reconstruction (3D-IIR). Methods: A phantom study and clinical study (31 patients) were performed with dual-energy CT (DECT). VMI at 40 keV was obtained and the images were reconstructed using filtered back projection (FBP), 50% adaptive statistical iterative reconstruction (ASiR), and 3D-IIR. We conducted subjective and objective evaluations of the image quality with each reconstruction technique. Results: The image contrast-to-noise ratio and image noise in both the clinical and phantom studies were significantly better with 3D-IIR than with 50% ASiR, and with 50% ASiR than with FBP (all, p < 0.05). The standard deviation and noise power spectra of the reconstructed images decreased in the order of 3D-IIR to 50% ASiR to FBP, while the modulation transfer function was maintained across the three reconstruction techniques. In most subjective evaluations in the clinical study, the image quality was significantly better with 3D-IIR than with 50% ASiR, and with 50% ASiR than with FBP (all, p < 0.001). Regarding the diagnostic acceptability, all images using 3D-IIR were evaluated as being fully or probably acceptable. Conclusions: The quality of VMI at 40 keV is improved by 3D-IIR, which allows the image noise to be reduced and structural details to be maintained. Advances in knowledge: The improvement of the image quality of VMI at 40 keV by 3D-IIR may increase the subjective acceptance in the clinical setting.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.