: Organ-based tube-current modulation has been used to reduce radiation exposure to speci c organs. However, there are no reports yet published on reducing radiation exposure in clinical cases. In this study, we assessed the reduction in radiation exposure to the mammary glands during thoracic computed tomography CT using X-CARE. In a phantom experiment, the use of X-CARE reduced radiation exposure at the midline of the precordial region by a maximum of 45.1 . In our corresponding clinical study, CT was performed using X-CARE in 15 patients, and without X-CARE in another 15. Compared to the non-X-CARE group, radiation exposure was reduced in the X-CARE group at the midline of the precordial region by 22.3 P 0.05 and at the medial sides of the right and left breasts by 16.8 and 14.2 , respectively P 0.05 for each . Analysis of the visual quality of CT images as well as CT values for the muscles of the chest walls showed no difference in either assessment between the X-CARE and non-Xcare groups P 0.05 . X-CARE thus reduced radiation exposure at the midline of the precordial region and allowed us to obtain consistent CT values without increasing noise. However, this study revealed increases in radiation exposure at the lateral sides of the breasts. It is conceivable that patients breasts were laterally displaced by gravity under the standard thoracic imaging conditions. Further studies that consider factors such as body size and adjustment of imaging conditions may be needed in the future.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.