During hemodialysis microembolic findings have been noted after the venous chamber (subclavian vein). The aim of this study was to evaluate if air could pass the venous chamber and, if so, if it passes the safety-system detector for air-infusion without triggering an alarm. Various in vitro dialysis settings were performed using regular dialysis devices. A dextran fluid was used instead of blood to avoid the risk of development of emboli. Optical visualization as well as recirculation and collection of eventual air into an intermediate bag were investigated. In addition, a specifically designed ultrasound monitor was placed after the venous air trap to measure the presence of eventual microbubbles. Speed of dialysis fluid was changed, as was the level of the fluid in the air trap. Thereby a fluid level was considered "high" if it was close to the top of the air trap and "low" if it was around the mid part of the air trap. By optical vision microbubbles were seen at the bottom of the air trap and could pass the air trap towards the venous line without alarming. During recirculation several mL of air were collected in an intermediate bag after the venous line. Ultrasound monitoring exhibited the presence of microbubbles of the size of approximately 5 microm upwards passing to the venous line in all runs performed. Amount of bubbles differed between devices and in general an increased fluid speed correlated significantly with the increased counts of microbubbles/min. No alarming of the detector occurred. A more concentrated fluid allowed higher counts/min when flow was increased to 600 mL/min. Data revealed that air passes the safety-sensor in the air trap without alarming. The presence of air increased in general with fluid speed and a lower fluid level in the air trap. Differences were present between devices. If this affects the patients has to be elucidated.
The infused and returning fluid from HD devices contains air microbubbles that enter the patient without triggering any alarms. These small emboli pass the lung and may cause ischaemic lesions in organs supported by the arterial circuit, such as the brain.
We previously showed, in vitro, that micro bubbles pass the air trap without inducing an alarm. The aim was to investigate if micro bubbles bypass the detector during hemodialysis (HD). During HD (40 patients, 47 HD sessions, 231 measurements), an ultrasound detector was fixed just after the venous air trap. Micro bubble size was measured in the range from 5 microm up to >42.5 microm. Blood flow was at a mean 346 mL/min (SD +/- 57). The mean of all micro bubbles per minute, without inducing an alarm, was at start 128 (range 0-769). Measurements revealed the presence of micro bubbles in all of the series and in 90% of the measurements. There was no difference between start and end of the same dialyses. There was a correlation between blood flow and extent of micro bubbles for the smaller sizes and the sum of all bubbles (r > or = 0.29, P < or = 0.026). Micro bubbles passed the air trap without alarming. Most bubbles were approximately 5 microm.
Venous air traps were tested in vitro with respect to presence of micro bubbles. Three types of venous air traps were measured (Bioline, Bioline GmbH, Luckenwalde, Germany; Gambro, Gambro AB, Lund, Sweden; Fresenius M.C., Fresenius Medical Care AG & Co. KGaA, Bad Homburg, Germany). Measurements (n = 10) were taken for each air trap, fluid flow (50-600 mL/min), and fluid level (high/low). A 1.5-MHz ultrasound probe was used with an analysis device. The probe was mounted on the outlet line downstream of the venous air trap. A semisynthetic fluid was used to resemble blood viscosity. Occurrences of micro bubbles, without inducing an alarm of the dialysis device, were detected in almost all measurements. The amount of bubbles increased with increasing flow. There were more bubbles with low fluid level compared with high level. The Bioline tubing released the least bubbles in high fluid level. At low level, the Gambro tubing showed the least bubbles at flows 50-400 mL/min, and the Fresenius M.C. tubing showed the least bubbles at flows 400-600 mL/min. High fluid level in the air trap reduced generation of micro bubbles compared to low level, as did lower fluid flow versus high flow. The design of the air trap was also of importance.
During hemodialysis (HD), blood that passes the dialysis device gets loaded with microbubbles (MB) of air that are returned to the patient without inducing an alarm. The aim with this study was to clarify if these signals are due to microembolies of air, clots, or artifacts, by histopathology of autopsy material of HD patients. These first results are from a patient on chronic HD. Due to pulmonary edema he was ultrafiltered. Within 30 minutes after the start, he suffered from a cardiac arrest and died. Autopsy verified the clinical findings. Microscopic investigation verified microembolies of air that were surrounded by fibrin in the lungs, brain, and heart. The study verified that MBs can enter the blood during HD and are trapped in the lungs. In addition, MBs pass the pulmonary capillaries and enter the arterial part of the body and are dispersed throughout the body. This can contribute to organ damage and be part of the poor prognoses seen in HD patients. Data support the importance to reduce MBs in the dialysis circuit.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.