Purpose
To evaluate heating of a redilatable stent for the treatment of aortic coarctation in neonates and small children in the new imaging modality magnetic particle imaging and established magnetic resonance imaging.
Materials and Methods
The cobalt-chromium stent (BabyStent, OSYPKA AG, Rheinfelden, Germany) has a stent design which allows for redilatation and adjustment of the diameter from 6 to 16 mm for a use in aortic coarctation. The stent loses its radial integrity while opening at predetermined breaking points at a diameter of 14 mm or 16 mm, respectively. We measured the temperature increase in the stent at different diameters during 7-min magnetic particle imaging and magnetic resonance imaging scans with fiber optic thermometers under static conditions surrounded by air. In magnetic particle imaging, stents with diameters from 6 to 16 mm were tested while in magnetic resonance imaging only stents with diameters of 6 mm and 14 mm were investigated exemplarily.
Result
In magnetic particle imaging, the measured temperature differences increased up to 4.7 K with growing diameters, whereas the opened stents with discontinuous struts at 14 and 16 mm showed only minimal heating of max. 0.5 K. In contrast to magnetic particle imaging, our measurements showed no heating of the stents during magnetic resonance imaging under identical conditions.
Conclusion
The BabyStent did show only slight heating in magnetic particle imaging and no detectable temperature increase in magnetic resonance imaging.
Purpose
Magnetic particle imaging (MPI) is a tomographic imaging modality with the potential for cardiovascular applications. In this context, the extent to which stents are heated should be estimated from safety perspective. Furthermore, the influence of the measurement parameters and stent distance to the isocenter of the MPI scanner on stent heating were evaluated.
Materials and Methods
Nine different endovascular stents and stent grafts were tested in polyvinyl-chloride tubes. The stents had diameters from 10 to 31 mm, lengths between 25 and 100 mm and were made from stainless steel, nitinol or cobalt-chromium. The temperature differences were recorded with fiber-optic thermometers. All measurements were performed in a preclinical commercial MPI scanner. The measurement parameters were varied (drive field strengths: 3, 6, 9, 12 mT and selection field gradients: 0, 1.25 and 2.5 T/m). Furthermore, measurements with different distances to the scanner’s isocenter were performed (100 to 0 mm).
Results
All stents showed heating (maximum 53.1 K, minimum 4.6 K). The stent diameter directly correlated with the temperature increase. The drive field strength influenced the heating of the stents, whereas the selection field gradient had no detectable impact. The heating of the stents decreased with increasing distance from the scanner’s isocenter and thus correlated with the loss of the scanner’s magnetic field.
Conclusion
Stents can cause potentially harmful heating in MPI. In addition to the stent diameter and design, the drive field strength and the distance to the MPI scanner’s isocenter must be kept in mind as influencing parameters.
Immune checkpoint inhibitor (ICI) therapy has revolutionized the treatment of several human malignancies, particularly metastatic skin cancer. However, immune-related myocarditis (irM), an immune-mediated adverse event (irAE), is often fatal. In the absence of a reliable biomarker, measurement of pre-ICI therapy serum troponin concentration has been proposed to identify patients at risk of developing irM, although real-world studies examining this strategy are lacking. Thus, we retrospectively analyzed the case records of all patients who commenced ICI therapy between January 2018 and December 2019 in a single university skin cancer center (n = 121) to (i) determine the incidence of irM, (ii) establish the frequency of pretreatment serum hsTnT elevations, and (iii) to establish whether this identified patients who subsequently developed irM. Only one patient developed irM, resulting in an overall incidence of 0.8%. Pretreatment hsTnT was measured in 47 patients and was elevated in 13 (28%). Elevated serum hsTnT concentrations were associated with chronic renal failure (p = 0.02) and diabetes (p < 0.0002). Pretreatment hsTnT was not elevated in the patient who developed fulminant irM. Pre-immunotherapy serum hsTnT concentrations were often asymptomatically elevated in patients with advanced skin cancer, none of whom subsequently developed irM during ICI therapy. However, large studies are required to assess the positive and negative predictive values of hsTnT for the development of irM. In the meantime, elevated hsTnT concentrations should be investigated before initiation of immunotherapy and closely monitored during early treatment cycles, where the risk of irM is greatest.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.