Conformal proton radiation therapy requires accurate prediction of the Bragg peak position. Protons may be more suitable than conventional x rays for this task since the relative electron density distribution can be measured directly with proton computed tomography ͑CT͒. However, proton CT has its own limitations, which need to be carefully studied before this technique can be introduced into routine clinical practice. In this work, we have used analytical relationships as well as the Monte Carlo simulation tool GEANT4 to study the principal resolution limits of proton CT. The noise level observed in proton CT images of a cylindrical water phantom with embedded tissueequivalent density inhomogeneities, which were generated based on GEANT4 simulations, compared well with predictions based on Tschalar's theory of energy loss straggling. The relationship between phantom thickness, initial energy, and the relative electron density resolution was systematically investigated to estimate the proton dose needed to obtain a given density resolution. We show that a reasonable density resolution can be achieved with a relatively small dose, which is comparable to or even lower than that of x-ray CT.
The new multistage design approach to WEPL measurements for proton CT and radiography has been prototyped and tested. The test results show that the design is competitive with much more expensive calorimeter and range-counter designs.
Proton computed tomography (pCT) has been proposed as an alternative to X-ray computed tomography (CT) for acquiring relative to water stopping power (RSP) maps used for proton treatment planning dose calculations. In parallel, it has been shown that dual energy X-ray CT (DECT) improves RSP accuracy when compared to conventional single energy X-ray CT. This study aimed at directly comparing the RSP accuracy of both modalities using phantoms scanned at an advanced prototype pCT scanner and a state-of-the-art DECT scanner. Two phantoms containing 13 tissue-mimicking inserts of known RSP were scanned at the pCT phase II prototype and a latest generation dual-source DECT scanner (Siemens SOMATOM Definition FORCE). RSP accuracy was compared by mean absolute percent error (MAPE) over all inserts. A highly realistic Monte Carlo (MC) simulation was used to gain insight on pCT image artifacts which degraded MAPE. MAPE was 0.55% for pCT and 0.67% for DECT. The realistic MC simulation agreed well with pCT measurements (MAPE = 0.69%).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.