SummaryWe compared the force of extraction for peripheral nerve catheters under three different situations in a porcine model using untunnelled, tunnelled and double-tunnelled catheters. Following insertion of the catheter into the porcine model, the catheters were either left untunnelled or a single or double tunnel was created for the catheter. The force required to displace the catheter by one centimetre was then measured in each of the three groups. The mean (SD) force required for displacement of the catheter was 0.23 (0.06) N for the untunnelled catheters, 1.16 (0.51) N for the single-tunnelled catheters, and 4.00 (1.70) N for the double-tunnelled catheters (p < 0.0001). Tunnelling a peripheral nerve catheter leads to a significant increase in the force required for dislodgement. This is increased further by introducing a second tunnel.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.