This work reports the results obtained from the comparison involving 7 laboratories in Brazil. This exercise is about the calibration service of surface contamination monitors. The monitor has been calibrated to BS ISO 7503-3, the calibration factor in terms of surface emission rate. The comparison was conducted by the Brazilian National Laboratory of Ionizing Radiation Metrology (LNMRI / IRD) from October 2018 to July 2019. The extensive sources used were 14C, 137Cs, 60Co, 90Sr/90Y, 36Cl and 241Am. The result of this proficiency test was excellent and proved the calibration capacity of the Brazilian network in the calibration service for surface contamination monitors.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.