Background
Recently many serological assays for detection of antibodies to SARS-COV-2 virus were introduced on the market. Aim of this study was to assess the diagnostic performance of an automated CLIA for quantitative detection of anti-SARS-CoV-2 IgM and IgG antibodies.
Methods
A total of 354 sera, 89 from consecutive patients diagnosed with COVID-19 (43 mild, 32 severe and 13 critical) and 265 from asymptomatic and negative on rRT-PCR testing healthcare workers, were evaluated for IgM and IgG anti-SARS-CoV-2 antibodies with MAGLUMI immunoasay.
Results
The overall sensitivity and specificity were 86.5% (95%CI: 77.6-92.8) and 98.5% (95%CI:96.2-99.6), respectively. PPV, PPN, LR+, LR- and OR were 95.1 (95%CI: 87.8-98.6), 95.6 (95%CI: 92.4-97.7), 57.3 (95%CI: 21.6-152.1), 7.3 (95%CI: 4.31-12.4) and 418.6 (95%CI: 131.2-1335.2), respectively. The levels of SARS-CoV-2 IgM and IgG antibodies were 1.22±1.2 AU/mL and 15.86±24.83 AU/mL, 2.86±2.4 AU/mL and 69.3±55.5 AU/mL, 2.47±1.33 AU/mL and 83.9±83.9 AU/mL in mild, severe and critical COVID-19 groups, respectively. A significant difference in antibody levels between mild and severe/critical subjects has been shown.
Conclusions
The CLIA assay showed good diagnostic performance and a significant association between antibody levels and severity of the disease was found.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.