Background: The measurement of renin activity is complicated by difficulties in the quantification of angiotensin 1 (Ang1), the product of the renin-catalyzed reaction. We report an HPLC-electrospray-tandem mass spectrometry (HPLC-ESI-MS/MS) method for the quantification of Ang1 as a measure of plasma renin activity (PRA).
Methods: After incubation (37 °C for 3 or 18 h), samples were prepared using C18 solid-phase extraction. [Val]5Ang1 was used as the internal standard (IS). Chromatography was performed on a C18 column, using 200 mL/L ammonium acetate buffer–800 mL/L methanol as the mobile phase. The flow rate was 150 μL/min, with a chromatographic run time of 5 min/sample. Mass spectrometric detection was in the positive ionization mode with selected reaction monitoring (Ang1 m/z 649.0→784.0; IS m/z 641.9→770.4).
Results: The assay was linear over the range 2.5–500 ng Ang1/mL, which corresponded to a limit of detection (signal-to-noise ratio of 3:1) of PRA of 0.14 ng Ang1 · mL−1 · h−1. The imprecision (CV) of the assay at PRA values of 26.1, 13.5, 3.2, and 0.78 ng Ang1 · mL−1 · h−1 was 7.0%, 7.0%, 15%, and 11%, respectively. Absolute recoveries were 92.3% (Ang1) and 87.4% (IS). Incubation times of 3 h vs 18 h in the PRA assay gave good agreement at PRA <2 ng Ang1 · mL−1 · h−1, but samples with a PRA of 2–5 ng Ang1 · mL−1 · h−1 gave lower PRA results after incubation for 18 h than after 3 h. We compared the HPLC-ESI-MS/MS assay and an RIA for the determination of PRA, with PRA incubation times of 3 h and 1.5 h, respectively. The mean PRA based on RIA of Ang1 was higher than that obtained using HPLC-ESI-MS/MS.
Conclusion: The HPLC-ESI-MS/MS method allows sensitive and specific measurement of PRA. The higher activities measured with the RIA method highlight its potential for overestimation of PRA.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.