This paper presents the results of the first characterization and experimental measurements of scatter dose at cardiologist's eyes for the only X-ray system that performs all paediatric Interventional cardiology procedures in Costa Rica. Entrance surface air kerma (ESAK) and the scatter dose values were measured on phantoms of 4-20 cm thicknesses of polymethyl methacrylate slabs. Image quality was evaluated using DICOM images of a test object Leeds TOR 18-FG, through the numerical parameters signal-to-noise ratio (SNR), high-contrast spatial resolution (HCSR) and figure of merit (FOM). When PMMA thickness is increased and during a move from low fluoroscopy to cine modes, ESAK, SNR, HCSR and FOM values range from 0.44 to 391.0 μGy fr-1; 2.8 to 14.89; 3.17 to 15.54 and 0.51 to 79.94, respectively. The highest scattered dose rates recorded during the simulations were 1.79 and 8.04 mSv h-1 for the high fluoroscopy and cine modes, respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.