We undertook a prospective, randomised study of 135 total knee arthroplasties to determine the most accurate and reliable technique for alignment of the tibial prosthesis. Tibial resection was guided by either intramedullary or extramedullary alignment jigs. Of the 135 knees, standardised postoperative radiographs suitable for assessment were available in 100. Correct tibial alignment was found in 85% of the intramedullary group compared with 65% of the extramedullary group (p = 0.019). We conclude that intramedullary guides are superior to extramedullary instruments for alignment of the tibial prosthesis.
We undertook a prospective, randomised study of 135 total knee arthroplasties to determine the most accurate and reliable technique for alignment of the tibial prosthesis. Tibial resection was guided by either intramedullary or extramedullary alignment jigs. Of the 135 knees, standardised postoperative radiographs suitable for assessment were available in 100. Correct tibial alignment was found in 85% of the intramedullary group compared with 65% of the extramedullary group (p = 0.019). We conclude that intramedullary guides are superior to extramedullary instruments for alignment of the tibial prosthesis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.