Background: Reliable serum creatinine measurements in glomerular filtration rate (GFR) estimation are critical to ongoing global public health efforts to increase the diagnosis and treatment of chronic kidney disease (CKD). We present an overview of the commonly used methods for the determination of serum creatinine, method limitations, and method performance in conjunction with the development of analytical performance criteria. Available resources for standardization of serum creatinine measurement are discussed, and recommendations for measurement improvement are given. Methods: The National Kidney Disease Education Program (NKDEP) Laboratory Working Group reviewed problems related to serum creatinine measurement for estimating GFR and prepared recommendations to standardize and improve creatinine measurement.
This is a PDF file of an article that has undergone enhancements after acceptance, such as the addition of a cover page and metadata, and formatting for readability, but it is not yet the definitive version of record. This version will undergo additional copyediting, typesetting and review before it is published in its final form, but we are providing this version to give early visibility of the article. Please note that, during the production process, errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
BACKGROUND
Methods from 7 manufacturers and 1 distributor for directly measuring HDL cholesterol (C) and LDL-C were evaluated for imprecision, trueness, total error, and specificity in nonfrozen serum samples.
METHODS
We performed each direct method according to the manufacturer’s instructions, using a Roche/Hitachi 917 analyzer, and compared the results with those obtained with reference measurement procedures for HDL-C and LDL-C. Imprecision was estimated for 35 runs performed with frozen pooled serum specimens and triplicate measurements on each individual sample. Sera from 37 individuals without disease and 138 with disease (primarily dyslipidemic and cardiovascular) were measured by each method. Trueness and total error were evaluated from the difference between the direct methods and reference measurement procedures. Specificity was evaluated from the dispersion in differences observed.
RESULTS
Imprecision data based on 4 frozen serum pools showed total CVs <3.7% for HDL-C and <4.4% for LDL-C. Bias for the nondiseased group ranged from −5.4% to 4.8% for HDL-C and from −6.8% to 1.1% for LDL-C, and for the diseased group from −8.6% to 8.8% for HDL-C and from −11.8% to 4.1% for LDL-C. Total error for the nondiseased group ranged from −13.4% to 13.6% for HDL-C and from −13.3% to 13.5% for LDL-C, and for the diseased group from −19.8% to 36.3% for HDL-C and from −26.6% to 31.9% for LDL-C.
CONCLUSIONS
Six of 8 HDL-C and 5 of 8 LDL-C direct methods met the National Cholesterol Education Program total error goals for nondiseased individuals. All the methods failed to meet these goals for diseased individuals, however, because of lack of specificity toward abnormal lipoproteins.
BACKGROUND: Urinary excretion of albumin indicates kidney damage and is recognized as a risk factor for progression of kidney disease and cardiovascular disease. The role of urinary albumin measurements has focused attention on the clinical need for accurate and clearly reported results. The National Kidney Disease Education Program and the IFCC convened a conference to assess the current state of preanalytical, analytical, and postanalytical issues affecting urine albumin measurements and to identify areas needing improvement.
CONTENT:The chemistry of albumin in urine is incompletely understood. Current guidelines recommend the use of the albumin/creatinine ratio (ACR) as a surrogate for the error-prone collection of timed urine samples. Although ACR results are affected by patient preparation and time of day of sample collection, neither is standardized. Considerable intermethod differences have been reported for both albumin and creatinine measurement, but trueness is unknown because there are no reference measurement procedures for albumin and no reference materials for either analyte in urine. The recommended reference intervals for the ACR do not take into account the large intergroup differences in creatinine excretion (e.g., related to differences in age, sex, and ethnicity) nor the continuous increase in risk related to albumin excretion.
DISCUSSION:Clinical needs have been identified for standardization of (a) urine collection methods, (b) urine albumin and creatinine measurements based on a complete reference system, (c) reporting of test results, and (d) reference intervals for the ACR.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.