Aluminum, yttrium, copper, and R-filters were evaluated with respect to effective beam energy, entrance radiation exposure to the skin, x-ray tube loading, and the quality of resultant radiographs. Since all filters yielded acceptable radiographs, a figure of merit was developed to determine which filter delivered the lowest entrance radiation exposure for the lowest increase in x-ray tube loading. The figure of merit indicated that the best results were obtained with the 0.1-mm Cu filter with 2 mm Al at 60-105 kVp and with the 0.2-mm Cu filter with 2 mm Al at 105-120 kVp. Cu filters produced radiographs with quality comparable to that of R- or Y filters. Cu filters are inexpensive alternatives to rare-earth filters and yielded a reduction in radiation exposure with less x-ray tube loading.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.