A personal dose equivalent comparison was carried out between the BEV and the PTB using a direct-reading electronic dosemeter system as a transfer instrument. This was the first comparison for the 'new' operational quantity, H p (10), for individual monitoring of strongly penetrating radiation and was performed in the preparatory stage of the European co-operation for Accreditation interlaboratory comparison IR4. For the BEV-PTB comparison described in this paper, the ISO water slab phantom and the ISO radiation qualities N-60 to N-300, S-Cs and S-Co were selected, whereby the axes of the radiation fields were perpendicular to the phantom surface. The results show very good agreement. In addition, the properties of the dosemeter system were investigated.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.