In this paper, a microfabricated inductive coil comprising of 125-turn coil windings and a MnZn-based magnetic core in a volume of 200 mm(3) is presented for the magnetic neural stimulation in a spinal cord. The coil winding with the parallel-linkage design instead of the typical serial-linkage one is proposed not only to provide better design flexibility to the current mode driving circuit but also to simplify the fabrication process of the 3-D inductive coil, which can further advance the coil miniaturization. Experimental results show the microcoil with a 1.5 A, 1 kHz square-wave current input can induce a voltages of ~220 μV on the conducting wire with an impedance of ~0.2 Ω @ 1 kHz, 1 mm separation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.