In hemodialysis, vascular access is usually achieved through an arteriovenous fistula, and a dislodged needle can cause varying degrees of injury to patients. In severe cases, the loss of blood can prove to be fatal. This study proposed a blood leakage detection device for patients during hemodialysis (HD). First, the device was tested on a phantom arm, and later in a clinical test on patients receiving HD. The thoughts of the patients and the nursing staff involved were surveyed before and after the introduction of the device. Analysis of the results indicated that the device achieved 100% and 98.9% accuracy rates on the phantom arm test and clinical test, respectively. The results suggested that patients believed the device could reduce their mental anxiety, and the nursing staff considered the device reliable and that it would enhance the quality of care. The proposed detection device can be extended to similar applications for preventing catheter dislodgement, and to improve patient safety and reduce the stress of clinical nursing staff.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.