Background:
An augmented reality tool allows visual tracking of real anatomical structures and superimposing virtual images, so it can be used for navigation of important structures during surgery.
Objectives:
The authors have developed a new occlusal splint-based optical navigation system for craniomaxillofacial surgery. In this study, the authors aim to measure the accuracy of the system and further analyze the main factors influencing precision.
Methods:
Ten beagle dogs were selected and a three-dimensional model was established through computed tomography scanning, dental model making, and laser scanning, and then registration was performed according to the tooth marking points. The bilateral mandibular osteotomy was performed on Beagle dogs under navigation system based on the occlusal splint. The left side was taken to compare the deviation between the preoperative plan and the surgical results, and the accuracy of distance and angle and the stability of the system were analyzed.
Results:
The average position deviation between the preoperative design and intraoperative navigation was: 0.01 ± 0.73 mm on the lateral height of the mandibular ramus, 0.26 ± 0.57 mm on the inner height of the mandibular ramus, and 0.20 ± 0.51 mm on the osteotomy length. The average angle deviation is 0.94° ± 1.38° on the angle between the mandibular osteotomy plane and ramus plane and 0.66° ± 0.97° on the angle of the retained mandibular angle. And most of the data showed good consistency.
Conclusions:
In summary, the accuracy of the system can meet clinical requirements and can be used as a useful tool to improve the accuracy of craniomaxillofacial surgery.
Pump-probe imaging was first used for quantitative analysis of melanin in dark circles’ skin to improve the ability to diagnose and treat dark circles on human skin. This study aimed to compare the distribution characteristics in melanin of lower eyelid skin tissues and to determine whether pump-probe imaging has potential for the classification of dark circles in vivo. Specimens obtained from 15 patients undergoing blepharoplasty were examined using pump-probe imaging. Furthermore, adjacent slices were respectively treated with hematoxylin–eosin (HE) and ferrous sulfate (FeSO4) staining for cross-references. Subsequently, the melanin content index (MCI) and mean fluorescence intensity (MFI) were quantitatively analyzed by the pump-probe imaging. The distribution of melanin granules in the pump-probe image and FeSO4 staining was consistent. Meanwhile, the tissues of the skin with dark circles and normal skin demonstrated significant differences in MCI and MFI. These differences can be used to distinguish the skin with dark circles from the normal skin. Pump-probe imaging could be used for the analysis of the microstructure and spectral characteristics of melanin granules in skin with dark circles. Significant differences were noted between the pigmented type of dark circles and the other two groups (normal skin and the vascular type of dark circles), while no significant differences were found between normal skin and the vascular type of dark circles.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.