Background
Sputum is commonly used for the diagnostic testing of pulmonary tuberculosis (PTB), but people living with HIV/AIDS (PLWHA) usually have little sputum. Moreover, the automated molecular test, Xpert MTB/RIF assay (Xpert), has a low sensitivity in PLWHA. We aimed to estimate the performance of Xpert Ultra on the detection of Mycobacterium tuberculosis (MTB) using bronchoalveolar lavage (BAL).
Methods
From February 5, 2018 to March 30, 2019, a total of 99 PLWHA with suspected PTB at the Third People’s Hospital of Shenzhen, China, were recruited. The information on demographics and medical history, blood MTB antigen-specific interferon gamma enzyme-linked immunospot assay (T-SPOT.TB), T lymphocyte subsets, and plasma HIV RNA load were collected. Computed tomography (CT) and flexible bronchoscopy were performed, and BAL and blood samples were collected. Testing of acid-fast bacilli (AFB), tuberculosis real-time fluorescence quantitative PCR (TBDNA), Ultra, Xpert, and MTB culture were conducted.
Results
Compared to BAL MTB culture for tuberculosis diagnosis, Ultra, Xpert, T-SPOT.TB, TBDNA and AFB smear had the sensitivity of 0.96 (24/25), 0.80 (20/25), 0.84 (21/25), 0.44 (11/25), and 0.12 (3/25), respectively; and the specificity of 0.92 (68/74), 0.96 (71/74), 0.93 (69/74), 0.96 (71/74), and 0.99 (73/74), respectively. Our study found that the sensitivity of Ultra was higher than that of culture and Xpert (AUC 0.92, 0.86 and 0.84, respectively). The results also indicated that PLWHA with CD4 <200 cells/mm
3
had reduced both sensitivity (from 1.00 and 0.86 to 0.94 and 0.78, respectively) and specificity (from 0.96 and 1.00 to 0.90 and 0.41, respectively) of Ultra and Xpert for the diagnosis of PTB.
Discussion
Our data supported an increased sensitivity of Ultra compared to that of Xpert on BAL samples of PLWHA, regardless of the CD4 counts and reference diagnosis standards.