Background: Drug-eluting stents (DESs) have been developed with thinner stent struts, and more biocompatible polymers and anti-proliferative drugs to improve the clinical performance. However, it remains unclear whether thinner struts are associated with favorable short-and long-term clinical outcomes such as target lesion revascularization (TLR), periprocedural myocardial infarction (PMI), and stent thrombosis (ST).Methods: We searched MEDLINE, Embase and other online sources for randomized controlled trials (RCTs) comparing clinical outcomes between a DES and other stent(s), with independent clinical event adjudication. We investigated stent-related events (TLR, PMI, and ST) in 5 years. Each outcome was analyzed with randomeffects meta-regression model against strut thickness, then adjusted for DES generation and patient and lesion characteristics.Results: We identified 49 RCTs enrolling 97,465 patients, of which strut thickness ranged from 60 to 140 μm. Incidences of 1-year TLR, PMI, and early ST were reduced with thinner stent struts, when adjusted for stent generation (adjusted relative risk[RR] per 10 μm increase 1.12 [95% CI 1.04-1.21], 1.15 [95% CI 1.05-1.26], and 1.15 [95% CI 1.06-1.25], respectively). Strut thickness was not independently associated with incidences of 5-year TLR, late and very late ST. In addition, early DESs contributed to a higher incidence of very late ).Conclusions: In this meta-regression analysis, a thinner strut thickness was associated with reduced incidences of early stent-related adverse events (1-year TLR, PMI, and early ST), but not with later events (5-year TLR, late ST, and very late ST).