We compared in vitro rates of hemolysis for a recently developed centrifugal pump with a conventional roller pump (10-10-00; Stöckert, Munich, Germany). Flow rates of 0.3 L/min and 1 L/min and a pressure of 200 mm Hg were chosen to simulate conditions during neonatal extracorporeal membrane oxygenation (ECMO). There was no significant difference in hemolysis rates between centrifugal and roller pumps (p = 0.57) nor between high and low flow (p = 0.86). The centrifugal pump caused no more blood trauma than the roller pump at the low-flow/high-pressure conditions required for neonatal ECMO. The Nikkiso pump is superior to roller pumps in size and priming volume (25 ml) and may permit development of a smaller and simpler ECMO system.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.