The ability to distinguish harmful solid cerebral emboli from gas bubbles intra-operatively has potential to direct interventions to reduce the risk of brain injury. In this in vitro study, two embolus discrimination techniques, dual-frequency (DF) and frequency modulation (FM) methods, are simultaneously compared to assess discrimination of potentially harmful large pieces of carotid plaque debris (0.5–1.55 mm) and thrombus-mimicking material (0.5–2 mm) from gas bubbles (0.01–2.5 mm). Detection of plaque and thrombus-mimic using the DF technique yielded disappointing results, with four out of five particles being misclassified (sensitivity: 18%; specificity: 89%). Although the FM method offered improved sensitivity, a higher number of false positives were observed (sensitivity: 72%; specificity: 50%). Optimum differentiation was achieved using the difference between peak embolus/blood ratio and mean embolus/blood ratio (sensitivity: 77%; specificity: 81%). We conclude that existing DF and FM techniques are unable to confidently distinguish large solid emboli from small gas bubbles (<50 μm).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.