A wireless intracranial pressure (ICP) monitoring system based on the air pressure sensing is proposed in this work. The proposed system is composed of an implantable ICP sensing device and a portable wireless data recorder. The ICP sensing device consists of an air pressure sensor, an ultra-thin air pouch for pressure sensing, and a low-power dedicated system-on-a-chip (SoC) for the data acquisition control and wireless transmission. The SoC consists of a power management unit, a wake-up controller, the sensor interface, a wireless transmitter, and the workflow control logic. The SoC is fabricated in 0.18 μm CMOS technology with a die area of 3.04 mm × 2 mm. Experimental results show that the prototype implantable ICP device has achieved a resolution of 0.2 mmHg and a battery lifetime of 1 week with a 3 V 50 mAh battery. The ICP device has been tested in the liquid environment. The nonlinearity error is less than ±0.4 mmHg for the full measurement range of -20 to +150 mmHg. Compared to the other implantable wireless ICP solutions in the literature, the proposed system alleviates the biocompatibility issue and increases the measurement accuracy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.