A novel stability indicating gradient reverse-phased high-performance liquid chromatographic method has been developed for the quantification of impurities of perindopril tert-butylamine (PER) in pharmaceutical dosage form. Separation of the active substance and its known (Impurities B, C, D, E, F) and unknown impurities was achieved on a BDS Hypersil C18 column (250 mm × 4.6 mm, 5 µm), thermostated at 70°C, using a mobile phase comprised of aqueous solution of sodium 1-heptanesulfonate adjusted to pH 2 with perchloric acid and acetonitrile. The flow rate was maintained at 1.5 mL min(-1), injection volume of 20 µL was utilized and detection of analytes was performed at 215 nm. The developed method was validated in accordance with current ICH Guidelines for all suggested parameters, including forced degradation studies and proved to be linear, accurate, precise and suitable for the impurity testing of PER, being subsequently applied during on-going stability studies of a newly developed generic formulation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.