High flow therapy works partly by washout of airway dead space, the volume of which has not been quantified in newborns. This observational study aimed to quantify airway dead space in infants and to compare efficacy of washout between high flow devices in three-dimensional (3D) printed airway models of infants weighing 2.5-3.8 kg. Nasopharyngeal airway dead space volume was 1.5-2.0 mL/kg in newborns. A single cannula device produced lower carbon dioxide (CO) levels than a dual cannula device (33.7, 31.2, 23.1, 15.9, 10.9 and 6.3 mm Hg vs 36.8, 35.5, 32.1, 26.8, 23.1 and 18.8 mm Hg at flow rates of 1, 2, 3, 4, 6 and 8 L/min, respectively; p<0.0001 at all flow rates). Airway pressure was 1 mm Hg at all flow rates with the single cannula but increased at higher flow rates with the dual cannula.Relative nasopharyngeal airway dead space volume is increased in newborns. In 3D-printed airway models, a single cannula high flow device produces improved CO washout with lower airway pressure.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.