The aim of this study is evaluation of dose distribution on radiation therapy planning system with the CT image of high-density material inserted phantom. Gammex 467 Tissue Characterization Phantom is used to acquire an image similar to the human tissues and insert a Titanium to generate metal artifact. The acquired images were reconstructed with Metal Artifact Reduction for Orthopedic Implants (O-MAR). By using the treatment planning system, the volume was analyzed and dose distribution was extracted. Photon dose distribution in linear accelerator was measured by the MapCHECK™ and compared with planned and measured dose distributions. In result of the comparative analysis, when artifact is generated by Titanium, The volume applied O-MAR was increased 6.8% to BR-12 Breast and 40.2% to LV 1 Liver. After O-MAR was used, Dose distribution was higher 1.4 to 1.6% than before. Consequently, The artifact caused by metal objects should be removed if possible, and after that used in the radiotherapy treatment plan can be considered to reduce errors.