“…A negligible change in diastolic wave speed was reported for ARF-based SWE (Δ0.04 m/s for ΔEDP of 14.7 mmHg (Pernot et al, 2011) and Δ0.5 kPa for ΔEDP of 5 mmHg (Pislaru et al, 2014b)), whereas the wave speed after MVC increased significantly with increased EDP (Δ4.9 m/s for ΔEDP of 21.7 mmHg: R=0.83 and p<0.001 (Bezy et al, 2020b)). Differences in shear wave excitation properties, experimental set-up and investigated pressure range partly explain the reported difference in the sensitivity of shear wave speed to pressure, but more importantly, the end-diastolic moment of the shear wave recording for ARF-based SWE might still be in the quasi-linear part of the stress-strain relationship whereas this might be different for the shear wave after MVC for which, again, the exact timing within the isovolumic contraction phase is unclear (Keijzer et al, 2020a;Pernot and Villemain, 2019).…”